Dimensions (D x W x H) Interior:
2400 x 2300 x 2270 mm
2400 x 3400 x 2270 mm
2400 x 4700 x 2270 mm
Unit Size:
Each
This item is not returnable. View return policy

Product Code. 17852755

Brand: BINDER™ 98000012

Only null left
Add to basket
Add to basket
This item is not returnable. View return policy

These walk-in-chambers offer a wide range of features and benefits to meet requirements in stability testing. Ideal test conditions are maintained in a wide range of temperatures, from 10°C to 50°C, with a relative humidity of 20% to 90% RH.

  • Relative humidity accuracy of ±2.5%
  • Temperature accuracy of ±1.5°C
  • Duct work for proper distribution of air
  • Highly insulated room panels with U value of only 0.21 W/m²K
  • LED lighting with motion detector
  • High quality room elements, made in Germany
  • Interior completely stainless steel
  • Exterior service accessibility
  • Room Elements: Galvanized sheet steel, polyester coated, white, similar to RAL 9010, interior surfaces in stainless steel
  • 4 inch/100 mm PU foam U value: 0.21 W/m²K
  • Floor: Floor overlay is chrome-nickel steel, pattern-rolled 5 WL
  • Surface load: 50,000 N/m²

Especificaciones

Chamber Material Stainless Steel
Lighting LED
Material Galvanized Steel
No. of Shelves 0
Plug Type EU, UK
Relative Humidity 20 to 90%
Temperature Range (Metric) 10°C to 50°C
Type Walk-in Chamber
Model WIC2-230V
Temperature Uniformity ±1.5 K at 40°C and 75% RH
Dimensions (D x W x H) Interior 2400 x 3400 x 2270 mm
Dimensions (L x W x H) 2600 x 3600 x 2470 mm
Weight (Metric) 981 kg
Voltage 200/230 V
Wattage 2.7 kW
Amperage 11.5 A
Frequency 50 Hz
Mostrar más Mostrar menos
Corrección del contenido de un producto

Proporcione sus comentarios sobre el contenido del producto rellenando el siguiente formulario.

Título del producto

Al hacer clic en Enviar, acepta que Fisher Scientific se ponga en contacto con usted en relación con los comentarios que ha proporcionado en este formulario. No compartiremos su información para ningún otro fin. Toda la información de contacto proporcionada se mantendrá de acuerdo con nuestra Política de Privacidad. Política de privacidad.

Gracias por ayudarnos a mejorar nuestra web. Su comentario ha sido enviado